|


The
measurement of UVA (or total UV) irradiance is often part of a daily
routine for indoor tanning salon operators. For reasons of practicality
and cost, the UVA measuring instrument preferred for this purpose
is generally the pocket-sized type.
We
all understand that the UVA reading plays an important part in deciding
when to replace tanning lamps. For financial reasons and customer
satisfaction, there should be the best possible assurance that the
salon owner can depend on the readout from his or her UV meter.
A UVA
meter can be a great tool when used properly; however, all too often
salon owners think it is an exact measurement for output, but instead
it is relative. If salon owners choose to rely on the UV meter then
they should be sure to follow the same format for measuring output
including:
- Take
an initial reading upon the installation of new lamps.
- Make
sure the "measuring conditions" always are kept the same. Measure
at the same location of your tanning unit, at the same distance,
observe the same electrical conditions for each single measurement.
- Make
sure that the acrylics and reflectors are clean. It is imperative
that acrylics are changed according to the manufacturer's specifications.
- Wait
with your measurement until your tanning unit is in a state of
"thermal equilibrium," which takes about 5 to 10 minutes after
turning it on.
If
these steps are carried out, measurement readings for comparative
purposes will be obtained. Pocket-sized meters are used to measure
a variety of UV bandwidths. Even if your meter only registers UVA
levels, it can be assumed that the UVB values will diminish by approximately
the same relative amount.
You
can check individual lamps at the acrylic, or take an average "body"
reading in the middle of the bed. Body position is generally considered
to be 25 cm above the bench, with meter pointed up at the closed
canopy.
Accuracy
Of Measurements
In general, the accuracy of UV readings on digital displays is sometimes
overestimated. Renowned universities consider an absolute accuracy
of between 5 percent and 10 percent as very good. Therefore, for
some low-cost UV meters that operate under much less controlled
conditions, considerably greater measuring errors could be expected.
For example, it is known that with some low-cost UV meters, the
readings for one and the same radiation source differ by up to 20
percent to 30 percent between identical products from the same manufacturer.
Some manufacturers of UV meters do provide calibration accuracy
of +/-5 to 10% reference NIST traceability, so it's important to
look for that specification to minimize error potential.
UVB
Measurement
It is more difficult for a hand-held meter to isolate the UVB bandwidth
(280-320 nm) from the total UV bandwidth (280-400 nm). Hence UVB
measurement can be more problematic. And since the primary purpose
of using a UV meter is to tell the salon owner when to replace his
or her lamps, the measurement of UVB values is not necessary for
that purpose because UVA and UVB irradiance drop by about the same
relative amount.
However,
a very important usage of a UVB meter is to check acrylic transmission.
Solarized, aged acrylic blocks a significant amount of UVB, while
allowing most UVA to transmit through it. Only a UVB meter that
has a selective 280-320 nm filter can display this problem. Another
usage for a UVB meter is to divide the reading by UVA value and
obtain an estimation of percent UVB.
Lamp
Output
When examining the output performance of a lamp, as a rule, about
20 percent to 25 percent of the electrical input will be emitted
as ultraviolet radiation. Yet, this only is true for new lamps that
are in optimal operating conditions--meaning that the electrical
conditions such as supplied voltage, lamp current and wattage meet
the specifications of the lamp manufacturer.
Ballasts
and starters--if used--also play an important role. The two major
factors responsible for the actual performance of a tanning unit
are the aging of the lamps over time and equipment characteristics
of the tanning unit.
Any
radiation source loses a certain degree of power the longer it is
in operation. For example, the reduction in power hardly is noticed
in the general lighting of private households. In general, incandescent
or fluorescent lamps are not replaced until they stop functioning.
Unfortunately,
this is not as simple with the UV lamps used for indoor tanning.
Tanning units are expected to meet certain requirements with respect
to their tanning effectiveness over a given period of time. With
increased use, they are no longer able to meet such expectations
because of a marked performance decline, even though they are still
in full working order from an electrical point of view. With tanning
lamps, the recommended "useful life" is considerably shorter than
the electrical life.
Normally,
manufacturers give recommendations on the "useful life" of their
lamps, but these recommendations only can be used as a guide because
there is no clear and official definition of the term "useful life."
Each manufacturer can make its own definition. Additionally, different
operating conditions, as well as equipment-related factors, have
in certain cases a considerable effect on the actual useful life
of a tanning lamp.
Although
both of these points are not very helpful in answering the question,
"When should lamps be replaced?", there are some guidelines which
help determine the right point in time.
Measuring
Different Types Of Lamps
Experience has proven that measuring different types of sunlamps
with a UVA or total UV meter usually proves problematic in practice.
The deviation of the UV meter from the actual measured value varies
from spectrum to spectrum. As a rule, UVA readings obtained on different
types of lamps are not comparable for effectiveness.
If
lamp (A) shows a reading of 20 mW/cm, and lamp (B) shows 15 mW/cm,
this does not generally mean that lamp (B) has a 25 percent lower
effectiveness. If there is a clear distinction between the spectral
characteristics of both lamps, it is perfectly possible that lamp
(B) has higher UVB with lower UVA… and tans in a lower session time.
That
is because the irradiance readings from UVA and UVB meters is unweighted
and does not correlate to the erythemal action spectrum (EAS). The
EAS weights 297 nm and below very high, and 298 out to 400 nm progressively
lower.
For
measuring different types of lamps' erythemal effectiveness, a MED/hr
meter must be used. If it's sensor response curve accurately follows
EAS weighting, it will yield Eeff in much the same manner as spectroradiometers
do for FDA determination of Te maximum timer interval of (4) MEDs
(minimum erythemal doseage).
Conclusion
This information is intended to emphasize the fact that low-cost,
pocket-size UV meters are suitable for a variety of distinct purposes,
depending on their specific response.
UVA
(or total UV): Relative measurements for lamp aging.
VB: Acrylic transmission and %UVB.
MED: Erythemally weighted effectiveness.
Additionally,
salon owners should not rely solely on customer response regarding
the output of the lamps, due to the many and varied skin types and
base tan levels. They should follow a typical lamp maintenance and
replacement routine in order to provide their customers with consistent
tans year-round.
NOTE:
Special thanks to Steve Mackin of Solartech Inc for his knowledgeable
input on metering techniques and parameters.
|